Radiometric Measurement
Radiometric Measurement
Radiometric Measurement
Radiometry is the measurement of energy or power in electromagnetic radiation fields or light. The average output power is the most common radiometric measurement since many light sources, including CW lasers and LEDs, emit output power that is constant over time. For a pulsed source, the pulse energy is typically the radiometric unit of measure, although the average output power can be given as well. Since sources can have different spatial distributions and divergences, other parameters may be needed to fully characterize their outputs. For this reason, this section will discuss the most commonly-encountered radiometric quantities for measuring power and energy (see Table 1). There can be considerable confusion regarding the nomenclature of radiometric terms, which can lead to measurement errors if not properly understood. The discussion below aims to adhere to the International Commission on Illumination (CIE) system, which fits well with the SI system of units. For a more complete description of radiometry, its history, and its concepts.
Quantity | Usual Symbol | Typical Units |
---|---|---|
Power | φ | W |
Energy | Qe | J |
Irradiance | E | W / m2 |
Fluence | F | J / m2 |
Radiant Intensity | Ir | W / sr |
Radiance | Lr | W / sr x m2 |
Table 1. Commonly used radiometric quantities.
The parameters in Table 1 are defined as follows:
- The average output power is defined as φ for a source with a continuous and stable output. For simplicity, φ is denoted as the power and is the radiometric quantity quoted most often.
- For a pulsed source, φ becomes a time-dependent quantity, i.e., φ(t), with a peak amplitude and a temporal shape (see Pulse Characterization). This amplitude is referred to as the peak output power or peak power. This pulsed quantity should not be confused with the average output power.
- The quantity Qe is the energy within a pulse of light. This quantity can be measured provided the temporal response of the sensor is fast enough; otherwise, it can be determined based on . and the repetition rate of the source.
- The irradiance (E) is essentially the power per unit area or power density. The terms exitance and intensity are often used synonymously with E since they have similar meanings and identical units of measure.
- The fluence (F) is associated with pulsed sources since it is the energy per unit area or energy density.
- The radiant intensity (Ir) and radiance (Lr) are derivatives of power and irradiance that also account for the divergence or spreading of the light source based on radiation into a solid angle. Lr is also important when determining light throughput in an optical system.
- If the word “spectral” is used before any radiometric quantity, it implies consideration of the wavelength dependence of the quantity. The measurement wavelength should be given when a spectral radiometric value is quoted.
In order to ensure that a sensor or detector (the two terms will be used interchangeably in this section) can accurately measure a radiometric quantity such as power or energy, it needs to be calibrated using a detection calibration standard provided by one of the national standards laboratories such as National Institute of Standards and Technology (NIST) or Physikalisch-Technische Bundesanstalt (PTB). Typically, the output of a source such as a spectrally-filtered lamp or a laser is measured by a NIST/PTB traceable sensor and this calibration is transferred to a master sensor which is, in turn, used to calibrate the sensor under test. Errors associated with each of these steps, along with any additional errors associated with spectral or temperature corrections, are combined to determine the total error associated with the calibrated sensor. The sensors should be periodically calibrated to ensure these errors remain reliable. Such absolute errors are related to the sensor’s ability to accurately measure the power or energy. This should not be confused with relative errors, which are related to the sensor’s precision. These errors are based on the noise characteristics of the detector.
All sensors that measure either power or energy from a laser or LED are typically described by a set of detector performance parameters. These parameters can be used as selection criteria when choosing the appropriate photodetector to characterize the source at hand. The three different types of sensors described in the following section each possess their own unique set of characteristics. These mainly result from the differences in their light-to-electrical signal conversion processes and are typically divided into thermal detectors (thermopile, pyroelectric) or photon detectors (photodiode). These detector performance parameters and how they differ for each type of sensor are the subjects of this section.
Spectral Responsivity
Responsivity is a measure of the transfer function between the input optical power and the output electrical signal from the detector. Thermal detectors convert a temperature change into a voltage; therefore, their responsivity is typically given in units of V/W. Photodiodes convert absorbed photons in a semiconductor into a current; therefore, the responsivity is given in A/W. Spectral responsivity describes how the detector will respond as a function of wavelength or photon energy. This spectral response can be dependent on a material’s transmission properties, e.g., a window in front of a sensor, or a material’s absorptive properties, e.g., a coating on a sensor or the sensor material itself. Thermal detectors respond to heat and so one watt delivered by a UV photon produces the same response as one watt delivered by an IR photon. Provided the material that is being heated has uniform absorption, the spectral responsivity is flat. The response for an ideal thermal detector is shown in Figure 1. Photon detectors produce at most a single response element, i.e., an electron-hole pair per incoming photon. The energy carried by individual photons is inversely proportional to the wavelength, and so, for the same input power, there are fewer UV photons per watt compared to IR photons. Accordingly, photon detector responsivity is significantly lower in the UV than in the IR (see Figure 1 for ideal response). Furthermore, due to the presence of the bandgap in semiconductors, only photons with energies above the bandgap energy (Eg) will be absorbed. Therefore, a photodiode will exhibit an abrupt increase in the responsivity for wavelengths just below the value associated with the bandgap (λg) while going to even shorter wavelengths will result in a decrease in responsivity like that described above. Figure 1 provides a representative spectral responsivity for a Si photodiode while actual responsivity curves for various photodiodes are given Photodiode Sensor Physics. Due to the strong wavelength-dependence of the responsivity for photodiodes, it is not uncommon for specification sheets to give a peak responsivity value at a wavelength and provide a relative spectral response curve.
Noise, Noise Equivalent Power, and Normalized Detectivity
Linearity and Dynamic Range
Temporal Response
Aperture
Damage
Types of Optical Sensors
The sensors discussed in this section are differentiated by the way in which they convert the incident light into an electrical signal. Thermal detectors work by converting the incident radiation into an increase in temperature. The temperature change is measured either by a voltage generated at the junction of dissimilar metals or by the pyroelectric effect. In either case, the heat-sensitive element is coated with a black material to enhance the absorption of the radiation. The material is designed to possess a large and uniform absorption leading to good responsivity over a wide spectral range. This is the major advantage of thermal detectors. As a result of the time required to effect a temperature change, thermal detectors are generally slow. In a photodiode, photons are absorbed in a semiconductor p-n junction giving rise to mobile charge carriers. The electrical conductivity of the material increases in proportion to the incident optical power. Applying an electric field to the junction causes the carriers to be transported, resulting in a measurable electric current in the circuit. The detectivity of photodiodes is typically much larger than that of their thermal counterparts and the mechanism for conversion can be quite fast. The main disadvantage of photodiodes is that their responsivity is strongly dependent on the wavelength of the incident light. Table 2 lists the typical characteristics of the three different sensor types.
Thermopiles | Pyroelectrics | Photodiodes |
---|---|---|
Measure power from mW up to 30 kW | Measure energy from sub µJ up to 40 J | Measure power from fW up to 30 W |
Response times of 1-3 seconds | Measure pulse widths up to 20 ms and repetition rates up to 25 kHz | Inexpensive and compact |
Linearity of ±1% | Duty cycle - pulse width up to 25% of time between pulses | Linearity of ±0.5% below saturation |
Relatively independent of beam size and position on detector | Accommodate beam sizes up to 96 mm diameter | Large dynamic range Ñ over 9 orders of magnitude in one sensor |
Broadband spectral response | Broadband spectral response | Large wavelength sensitivity Ñ must specify exact wavelength |
High laser damage threshold | Can measure low level CW powers with chopper | Can measure sub µJ energies with reduced dynamic range |
Table 2. Typical characteristics of various sensor types.
Related Topics
Optical Sensors
Beam Characterization